jakemcivor commented Mar 6, 2024

Periodically during use, the third light will light up and stay lit up. Puffs will generate right clicks rather than left clicks, no mouse movements. Seems to resolve itself after a couple minutes and rapidly dump any actions that were captured when it was stuck (e.g., series of left / right clicks)

@jakemcivor jakemcivor self-assigned this Mar 12, 2024
@jakemcivor jakemcivor added the bug Something isn't working label Mar 12, 2024
@jakemcivor
Copy link
Member Author

This issue seems to be related or made worse by having a shortened mouthpiece. We are investigating the possibility of a saturated filter contributing to the issue as well.

@jakemcivor
Copy link
Member Author

jakemcivor commented Aug 25, 2024

Got a report that this is still occurring occasionally. The user rarely uses the puff function, but this behaviour still occurs. Changing the filter or modifying the mouthpiece length doesn't seem to resolve the issue.

To resolve, the user has to wait up to about 5 minutes, or needs to physically unplug the device.